NUMERICAL ANALYSIS OF THE EFFECT OF IMPLANT GEOMETRY TO STRESS DISTRIBUTIONS OF DENTAL IMPLANT SYSTEM

Volume: 18 Number: 1 February 11, 2015
EN

NUMERICAL ANALYSIS OF THE EFFECT OF IMPLANT GEOMETRY TO STRESS DISTRIBUTIONS OF DENTAL IMPLANT SYSTEM

Abstract

ABSTRACT

Purpose: The success of dental implants is related to the quality, quantity of local bones, implant design and surgical technique. Implant diameter and length are accepted as key factors. Present work focuses to investigate the effect of titanium implant geometry to stress distributions in implant system.

Materials and Methods: For this purpose three different implant models which are currently being used in clinical cases constructed by using ANSYS Workbench 12.1. The stress distributions on components of implant system  under static loadings were analyzed for all models.

Results: The maximum stress values that occurred in all components happen in the case of loading in which the Nucleoss T-4 implant is used, but the occurred lowest stress values happen in the case of Fı loading in which Nobel Active  implant is used. In all models, the maximum tensions have occurred in the neck region of the implants.

Conclusion: The crested loss of bones in the neck region of the implants lessen the long-term survival rate of implants.The lengt and the size of the implant are the two important factors in the stress distribution. 

Keywords

References

  1. Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg 1969;3:81-100.
  2. Buser D, Mericske-Stern R, Bernard JP, Behneke A, Behneke N, Hirt HP, et al. Long-term evaluation of non-submerged ITI implants. Part 1: 8-year life table analysis of a prospective multi-center study with 2359 implants. Clin Oral Implants Res 1997;8:161-172.
  3. Bozkaya D, Muftu S, Muftu A. Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. J Prosthet Dent 2004;92:523-530.
  4. Petrie CS, Williams JL. Comparative evaluation of implant designs: influence of diameter, length, and taper on strains in the alveolar crest. A three-dimensional finite-element analysis. Clin Oral Implants Res 2005;16:486-494.
  5. Acar A, İnan Ö. İmplant destekli protezlerde okluzyon. Cumhuriyet Üniversitesi Dişhekimliği Fakültesi Dergisi 2001;4,1:52-56. 6. Chun HJ, Cheong SY, Han JH, Heo SJ, Chung JP, Rhyu IC, et al. Evaluation of design parameters of osseointegrated dental implants using finite element analysis. J Oral Rehabil 2002;29:565-574.
  6. Tada S, Stegaroiu R, Kitamura E, Miyakawa O, Kusakari H. Influence of implant design and bone quality on stress/strain distribution in bone around implants: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants 2003;18:357-368.
  7. Anitua E, Tapia R, Luzuriaga F, Orive G. Influence of implant length, diameter, and geometry on stress distribution: a finite element analysis. Int J Periodontics Restorative Dent 2010;30:89-95.
  8. Teixeira ER, Sato Y, Akagawa Y, Shindoi N. A comparative evaluation of mandibular finite element models with different lengths and elements for implant biomechanics. J Oral Rehabil 1998;25:299-303.

Details

Primary Language

English

Subjects

-

Journal Section

-

Publication Date

February 11, 2015

Submission Date

June 25, 2014

Acceptance Date

-

Published in Issue

Year 1970 Volume: 18 Number: 1

EndNote
Topkaya T, Solmaz M, Dündar S, Eltas A (February 1, 2015) NUMERICAL ANALYSIS OF THE EFFECT OF IMPLANT GEOMETRY TO STRESS DISTRIBUTIONS OF DENTAL IMPLANT SYSTEM. Cumhuriyet Dental Journal 18 1 17–24.

Cited By

Cumhuriyet Dental Journal (Cumhuriyet Dent J, CDJ) is the official publication of Cumhuriyet University Faculty of Dentistry. CDJ is an international journal dedicated to the latest advancement of dentistry. The aim of this journal is to provide a platform for scientists and academicians all over the world to promote, share, and discuss various new issues and developments in different areas of dentistry. First issue of the Journal of Cumhuriyet University Faculty of Dentistry was published in 1998. In 2010, journal's name was changed as Cumhuriyet Dental Journal. Journal’s publication language is English.


CDJ accepts articles in English. Submitting a paper to CDJ is free of charges. In addition, CDJ has not have article processing charges.

Frequency: Four times a year (March, June, September, and December)

IMPORTANT NOTICE

All users of Cumhuriyet Dental Journal should visit to their user's home page through the "https://dergipark.org.tr/tr/user" " or "https://dergipark.org.tr/en/user" links to update their incomplete information shown in blue or yellow warnings and update their e-mail addresses and information to the DergiPark system. Otherwise, the e-mails from the journal will not be seen or fall into the SPAM folder. Please fill in all missing part in the relevant field.

Please visit journal's AUTHOR GUIDELINE to see revised policy and submission rules to be held since 2020.