COMPARISON OF CYCLIC FATIGUE RESISTANCE OF ONE CURVE, HYFLEX EDM, WAVEONE GOLD AND RECIPROC BLUE NICKELTITANIUM ROTARY FILES AT INTRA-CANAL TEMPERATURE
Abstract
Objectives: To compare the cyclic fatigue resistances of One Curve (OC), Hyflex EDM (HEDM), WaveOne Gold (WOG), Reciproc Blue (RPC Blue), and nickel-titanium rotary files at intra-canal temperature (35°C). Materials and Methods: Twenty OC (25/.06), 20 HEDM (25/.08), 20 WOG (25/.07) and 20 RPC Blue (25/.08) files were tested for cyclic fatigue at intra-canal temperature (35°C). All the instruments were rotated in artificial which were made of stainless steel with an 60° angle of curvature and a radii of curvatures of 5 mm until fracture occurred, and the time to fracture was recorded in seconds using a digital chronometer and the number of cycles to fracture (NCF) for each file was calculated. The Kruskal-Wallis test was performed to statistically analyze the data by using SPSS 21.0 software. The statistical significance level was set at p<.05.
Results: NCF values revealed that the HEDM had the highest cyclic fatigue resistance at intracanal temperature (p<.05) and followed by WOG, RPC Blue and OC. The lowest NCF values were determined in the OC group and statistically significant difference was observed between the OC group the other groups (p<.05). No statistically significant difference in the mean length of the fractured fragments in curvature was evident for the instruments (p> .05).
Conclusion: Within the limitations of the present in-vitro study, HEDM instrument resisted static cyclic fatigue significantly more than RPC Blue, WOG and OC instruments.
Keywords
References
- 1) Ankrum MT, Hartwell GR, Truitt JE. K3 Endo, ProTaper, and ProFile systems: breakage and distortion in severely curved roots of molars. J Endod 2004 (30): 234–7.2) Shen Y, Cheung GSP, Bian Z, Peng B. Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 2006 (32):61–5.3) Martin B, Zelada G, Varela P, et al. Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J 2003 (36):262–6.4) Kitchens GG, Liewehr FR, Moon PC. The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod 2007 (33):52–4.5) Bui TB, Mitchell JC, Baumgartner JC. Effect of electropolishing ProFile nickel–titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and cutting efficiency. J Endod 2008 (34):190–3.6) Peters OA, Gluskin AK, Weiss RA, Han JT. An in vitro assessment of the physical properties of novel Hyflex nickel–titanium rotary instruments. Int Endod J 2012 (45):1027-347) Gündoğar M, Özyürek. Cyclic Fatigue Resistance of OneShape, HyFlex EDM, WaveOne Gold,and Reciproc Blue Nickel-titanium Instruments. J Endod 2017 (43): 1192-68) Pedulla E, Lo Savio F, Boninelli S, Plotino G, Grande NM, La Rosa G, Rapisarda E. Torsional and Cyclic Fatigue Resistance of a New Nickel-Titanium Instrument Manufactured by Electrical Discharge Machining. J Endod 2016 (42):156-99) Yılmaz K, Uslu G, Özyürek T. Effect of multiple autoclave cycles on the surface roughness of HyFlex CM and HyFlex EDM files: an atomic force microscopy study. Clin Oral Investig 2018; Epub ahead of print10) WaveOne Gold brochure. Available at: https://www.dentsply.com/content/dam/ dentsply/pim/manufacturer/Endodontics/Obturation/Gutta_Percha_Points/WaveOne_Gold_Gutta_Percha_Points/W1G_Brochure_EN.pdf. Accessed July 10, 2017.11) Özyürek T. Cyclic Fatigue Resistance of Reciproc, WaveOne, and WaveOne Gold Nickel-Titanium Instruments. J Endod 2016 (42):1536-912) Özyürek T, Uslu G, Yılmaz K, Gündoğar M. Effect of Glide Path Creating on Cyclic Fatigue Resistance of Reciproc and Reciproc Blue Nickel-titanium Files: A Laboratory Study. J Endod 2018 (44):1033-713) Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. Int Endod J 2012 (45): 614-814) de Vasconcelos RA, Murphy S, Carvalho CA, Govindjee RG, Govindjee S, Peters OA. Evidence for Reduced Fatigue Resistance of Contemporary Rotary Instruments Exposed to Body Temperature. J Endod 2016 (42): 782-715) Shahabinejad H, Ghassemi A, Pishbin L, Shahravan A. Success of ultrasonic technique in removing fractured rotary nickel-titanium endodontic instruments from root canals and its effect on the required force for root fracture. J Endod 2013 (39):824–8.16) Cheung G, Peng B, Bian Z, et al. Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J 2005 (38):802–9.17) Inan U, Gonulol N. Deformation and fracture of Mtwo rotary nickel-titanium instrumentsafter clinical use. J Endod 2009 (35):1396–9.18) Castello-Escriva R, Alegre-Domingo T, Faus-Matoses V, et al. In vitro comparison of cyclic fatigue resistance of ProTaper, WaveOne and Twisted Files. J Endod 2012 (38): 1521–4.19) Yared G. Canal preparation using only one Ni-Ti rotary instrument: preliminary observations. Int Endod J 2008 (41):339–44.20)Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009 (35):1469-7621) Yılmaz K, Uslu G, Gündoğar M, Özyürek T, Grande NM, Plotino G. Cyclic fatigue resistances of several nickel-titanium glide path rotary and reciprocating instruments at body temperature. Int Endod J 2018 (51): 924-3022) de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P. İn Vivo Intracanal Temperature Evolution during Endodontic Treatment after the Injection of Room Temperature or Preheated Sodium Hypochlorite. J Endod 2015 (41): 1112-523) Özyürek T, Gündoğar M, Uslu G, Yılmaz K, Staffoli S,Gramde NM, Plotino G, Polimeni A. Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals. Odontology 2018: Epub Ahead of Print
Details
Primary Language
English
Subjects
Health Care Administration
Journal Section
Research Article
Authors
Koray Yılmaz
*
0000-0001-6096-7385
Türkiye
Taha Özyürek
0000-0003-3299-3361
Türkiye
Gülşah Uslu
0000-0003-3176-1251
Türkiye
Publication Date
March 13, 2019
Submission Date
November 1, 2018
Acceptance Date
January 4, 2019
Published in Issue
Year 2019 Volume: 22 Number: 1
Cited By
Impact of Peracetic Acid on the Dynamic Cyclic Fatigue of Heat-Treated Nickel-Titanium Rotary Endodontic Instrument
International Journal of Dentistry
https://doi.org/10.1155/2021/6676005Fracture of endodontic instruments - Part 1: Literature review on factors that influence instrument breakage
South African Dental Journal
https://doi.org/10.17159/2519-0105/2020/v75no10a4ÜÇ FARKLI NİKEL-TİTANYUM DÖNER ALET SİSTEMİNİN VÜCUT ISISINDAKİ DÖNGÜSEL YORGUNLUK DİRENÇLERİNİN KARŞILAŞTIRMALI OLARAK DEĞERLENDİRİLMESİ
Atatürk Üniversitesi Diş Hekimliği Fakültesi Dergisi
https://doi.org/10.17567/ataunidfd.1009686A comparison of failure rates and canal preparation times between WaveOne Gold and One Curve file systems with and without glide path preparation in simulated canals
South African Dental Journal
https://doi.org/10.17159/10.17159/2519-0105/2021/v76no5a6A single-file endodontic treatment: A promising endodontic concept
Vojnosanitetski pregled
https://doi.org/10.2298/VSP210524078Z