İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi

Volume: 12 Number: 2 October 30, 2011
  • Hale Cimilli
  • Seda Aydemir
  • Nevin Kartal
EN TR

İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi

Abstract

Purpose: The aim of this ex vivo study was to evaluate the accuracy of two apex locators: Propex (Dentsply, Ballaigues Switzerland) and RomiAPEX™ (Romidan, Tel Aviv, Israel).

Materials and methods: The accuracy of the measurement of two different electronic apex loctors were compared in root canal length determination by using forty extracted human teeth . The actual canal length was determined by introducing a size 15 K-file in the canal until the file tip became visible at the apical foramen under X5 magnification by using an operating microscope. A caliper to the nearest 0.5 mm were used during measurements. The teeth were embedded in acrylic as the apical one thirds and crowns were apparent. The measurements were taken as the apical one thirds embedded in alginate by using two electronic apex locators. All measurements were taken two times.

Results: The accuracy of the electronical working length measurements were found for Propex as 97.5% ; for RomiAPEX™ D-30 as 95%. No significant differences were found between actual root canal lengths and the measurement values of apex locators statistically (p>0,05).

Conclusion: High accuracy were confirmed for two apex locators during the measurements of root canal lengths.

ÖZET

Amaç: Bu ex vivo çalışmada Propex (Dentsply, Ballaigues, İsviçre) ve RomiAPEX™ D-30 (Romidan, Tel Aviv, İsrail) isimli iki farklı apeks bulucunun klinik doğruluğunun değerlendirilmesi amaçlanmıtır.

Gereç ve Yöntem: Bu çalımada kırk adet üst kesici diş kullanıldı. İki farklı apeks bulucu (Propex ve RomiAPEX™ D-30) kullanıldı. Dişlerin gerçek kanal boyları operasyon mikroskobunda X5 büyütmede #15 K tipi eğe ucu apikal foramenden görülene kadar ilerletilerek ölçüldü. Ölçümler sırasında 0.5 mm aralıklı kaliper kullanıldı. Dişler kronları ve apikal 1/3’leri açıkta kalacak şekilde akriliğe gömüldü. Ardından dilerin apikal 1/3’lük kısımları aljinata gömüldü ve iki apeks bulucuyla ölçümler yapıldı. Tüm ölçümler ikişer defa tekrarlandı.

Bulgular: Dişlerin elektronik kök kanal boyu ölçüm değerleri ortalaması gerçek kanal boyuyla karşılatırıldığı zaman ± 0.5 mm aralığında Propex için ölçüm değeri doğruluk oranı %97.5; RomiAPEX™ D-30 için %95 olarak hesaplandı. Gerçek kök kanal boyu ile elektronik apeks bulucuların ölçüm değerleri arasında istatistiksel olarak anlamlı bir farklılık bulunamadı (p>0,05).

Sonuç: Her iki apeks bulucunun da kök kanal boyu ölçümünde yüksek oranda doğruluk gösterdiği tespit edildi.

Keywords

References

  1. Kuttler Y (1955) Microscopic investigation of root apexes. JADA 50, 544-52.
  2. Ricucci D, Langeland K.Apical limit of root canal instrumentation and obturation, part 2. A histological study. Int Endod J 1998; 31:394-409.
  3. Dummer PMH, McGlinn JH, Rees DG. The position and topography of the canal constriction and apical foramen. Int Endod J ; 17, 192-8. Suzuki K (1942) Instruments, materials and devices. In Cohen S,
  4. Hargreaves KM, eds. Pathways of the Pulp, 9th edn. St Louis,MO: Mosby, pp.233-289. Sunada I. New method for measuring the length of the root canal. J Dent Res 1962; 41:375-87.
  5. Romidan Ltd. Kiryat-Ono, Israel. 2006, Instruction Guidelines.
  6. Bernardes BA, Duarte A.H, Vasconcelos B.C. Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and RomiAPEX D-30. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007; 104:e91-94.
  7. Plotino G, Grande NM, Brigante L, Lesti B, Soma F. Ex vivo accuracy of three electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator and Propex. Int Endod J 2006; , 408-414.
  8. Tinaz AC, Alacam T, Topuz O. A simple model to demonstrate the electronic apex locator. Int Endod J 2002; 35, 940-5.

Details

Primary Language

English

Subjects

-

Journal Section

-

Authors

Hale Cimilli

Seda Aydemir

Nevin Kartal

Publication Date

October 30, 2011

Submission Date

October 30, 2011

Acceptance Date

-

Published in Issue

Year 1970 Volume: 12 Number: 2

EndNote
Cimilli H, Aydemir S, Kartal N (October 1, 2011) İki farklı apeks bulucunun kök kanal boyu ölçümünde doğruluğunun ex vivo olarak değerlendirilmesi. Cumhuriyet Dental Journal 12 2 125–128.

Cumhuriyet Dental Journal (Cumhuriyet Dent J, CDJ) is the official publication of Cumhuriyet University Faculty of Dentistry. CDJ is an international journal dedicated to the latest advancement of dentistry. The aim of this journal is to provide a platform for scientists and academicians all over the world to promote, share, and discuss various new issues and developments in different areas of dentistry. First issue of the Journal of Cumhuriyet University Faculty of Dentistry was published in 1998. In 2010, journal's name was changed as Cumhuriyet Dental Journal. Journal’s publication language is English.


CDJ accepts articles in English. Submitting a paper to CDJ is free of charges. In addition, CDJ has not have article processing charges.

Frequency: Four times a year (March, June, September, and December)

IMPORTANT NOTICE

All users of Cumhuriyet Dental Journal should visit to their user's home page through the "https://dergipark.org.tr/tr/user" " or "https://dergipark.org.tr/en/user" links to update their incomplete information shown in blue or yellow warnings and update their e-mail addresses and information to the DergiPark system. Otherwise, the e-mails from the journal will not be seen or fall into the SPAM folder. Please fill in all missing part in the relevant field.

Please visit journal's AUTHOR GUIDELINE to see revised policy and submission rules to be held since 2020.